Test Details

Mnemonic

INFBG

Ordering Code

3684050

CPT-4 Code(s)

86710

Specimen Information

Collect

Serum separator tube (SST)

Specimen Preparation

Centrifuge, separate serum from cells within 2 hours and send 1.0 mL serum in a screw capped plastic vial.

Alternate Specimen

Serum: Red top

Minimum Volume

0.05 mL

Transport Temperature

Refrigerated

Rejection Criteria

Plasma, hemolyzed, lipemic, icteric, heat inactivated or turbid specimens

Stability

Room temperature: 2 days
Refrigerated: 14 days
Frozen: 1 year

Methodology

Semi-quantitative Enzyme Linked Immunosorbent Assay
more

Reference Range

Components Reference Interval
Influenza B Virus Antibody, IGG ≤0.89 IV:
Negative- No significant level of
influenza B virus IgG antibody detected.
0.90 - 1.10 IV: Equivocal- Questionable
presence of influenza B virus IgG antibody
detected. Repeat testing in 10 - 14 days
may be helpful.
≥1.11 IV: Positive- IgG antibodies to
influenza B virus detected, which
may suggest current or past infection.
more

Performing Laboratory

ARUP Reference Laboratory
more

New York DOH Approval Status

Yes
more