Updated Test
Effective Date
None Selected
Included in Report
Updated Test
Update to alternate specimen and example report on website.
Alternate Specimen: One oropharyngeal swab or NP/OP sent frozen in viral transport media.
Nasal swab sent frozen in viral transport media.
Our internal studies show that Phosphate Buffered Saline (PBS) and sterile saline do not interfere with the analytical performance of the COVID-19 assay. Liquid Amies buffer may decrease the analytical sensitivity of the assay and should be used only when other transport media are not available.
Bronchoalveolar lavage/wash in sterile, leak-proof container.
CVPCR
3000878
87635